Checkline TI-MINIMAX User Manual

Page 62

Advertising
background image

Dakota Ultrasonics

56

10) Repeat steps

8 & 9 until the PHYSICAL LEN value is correctly

displayed.

11) Press the

OK key to return to the One Point Zero List, or ESC to

cancel entering the

PHYSICAL LEN value.

12) Use the

UP and DOWN arrow keys to scroll through the One Point

Zero List Items until VELOCITY is highlighted.

13) Press the

ENTER key to display the Digits Edit Box.

Note: Enter the velocity value of the bolt that was documented in the

previous section. Reminder: The value was written down and scribed on

the side of the bolt. It is the velocity used for the probe zero bolt.

14) Press the

UP and DOWN arrow keys to scroll the highlighted value.

15) Press the

LEFT and RIGHT arrow keys to scroll the digit locations.

16) Repeat steps

14 & 15 until the VELOCITY value is correctly displayed.

17) Press the

OK key to return to the One Point Zero List, or ESC to

cancel entering the

VELOCITY value.

18) Use the

UP and DOWN arrow keys to scroll through the One Point

Zero List Items until TEMP COEF is highlighted.

Advertising