MINI-MAX Ultrasonic Bolt Tension Monitor

13) Press the

MENU key multiple times to tab right and the ESC key

multiple times to tab left until the

AUTO menu is highlighted and

displaying the submenu items.

14) Use the and

DOWN arrow keys to scroll through the sub menu items

until

AUTO SET is highlighted.

Note: The Auto Set feature can also be activated by simply pressing the

Auto Set key on the keypad.

15) Apply a drop of couplant to the cal standard or transducer, and attach it

to one end of the cal standard. Rotate the transducer clockwise and

counter clockwise applying a small amount of pressure to eliminate any

excess couplant between the transducer and cal standard. Be sure to

always place the transducer in the same location. This will help to

eliminate any potential measurement errors caused by changing the

sound path.

16) Press the

ENTER key and the Ultrasonic Length of the bolt will be

displayed.

17) Press the

MEAS key once to activate the measure menu items. Press

the

MEAS key multiple times to move right and the ESC key multiple

times to move left, until the

LOC cell is highlighted.

18) Press the

ENTER key to display the Group View Box.

19) Press the

UP, DOWN, LEFT, and RIGHT arrow keys to scroll the target

cell cursor to the desired storage location.