Wavetronix Command Collector (CMD-DCx) - User Guide User Manual

Page 62

Advertising
background image

CHAPTER 5 • SENSOR CONFIGURATION PAGE

61

Figure 5.11 – Sensor Command Tab

The following is a list of basic commands:

˽

Update Configuration – Attempts to download the current sensor configuration from

the sensor. Although Collector will automatically update the sensor configuration on a

timed interval, a manual update configuration command can be sent at any time.

˽

Sync Sensor Time – Attempts to set the onboard sensor time equal to the current

server time. Collector will automatically synchronize the sensor time on a timed inter-

val, although a manual sync time command can be sent at any time.

˽

Set Sensor Interval – Attempts to set the sensor interval to the given value in seconds.

Acceptable sensor intervals depend on the individual sensor firmware; consult the sen-

sor documentation for any limitations on sensor intervals. Changing the sensor inter-

val in this tab does not affect the collection interval that Collector uses to communicate

with the sensor; the collection interval must be changed separately.

Collector also supports any number of custom commands that a particular sensor brand

(and driver) supports. Any additional commands supported by the driver will be listed in

the Custom Command section.

Follow the steps below to run a custom command:
1 Select the Custom Command radio button.

2 Select the command name from the Name drop-down menu.

3 If necessary, fill in any additional parameters required by the command in the Proper-

ties table.

4 Click the Run Command button.

Note

After clicking the

Run Command button, a loading image will appear until the result

is displayed or the timeout is reached. It is possible for the timeout to hit before the
command completes successfully.

Advertising