Qc-300 series user’s guide – HEIDENHAIN ND 1300 OED and Crosshair Systems User Manual
Page 198

5 Enter the MZ Offset value noted earlier in step 2.
6 Zero the DRO for the axis being calibrated at the part zero position of the
standard and enter zeros into the Standard and Observed data fields for Station
zero.
7 Measure the standard, increment-
ing the Station number and entering
the Standard (certified) and Observed
(measured) values into the correspond-
ing data fields for each segment.
This example shows measurements
across 150 mm on the X-axis, and the
SLEC screens for entering the six seg-
ments of standard and observed data to
provide error correction.
Reminder
Segments are straight lines on the
deviation graph. Only the beginning
and end points of each straight line
segments must be entered into SLEC
data fields.
8 Touch the Enabled choice field to toggle it On.
Station 1 of 6
Station 2 of 6
Station 3 of 6
Station 4 of 6
Station 5 of 6
Station 6 of 6