9.2 Calibration Using Mag-Prover’s Built-In Signal Source

9.2.1 0 % Flow Rate Calibration (Zero Calibration)

Using Mag-Prover’s internal calibration circuit, 0% flow rate calibration (hereafter called zero
calibration) can be performed.

●Zero point check / calibration

Switch operation

Display example

Description

Select "CAL 0" from the setting item
selection menu.

When the calibration screen is selected, the
internal simulation circuit starts working and
0%

value using the internal simulation

signal appears.
Then push and hold longer.
* Pushing returns you to the setting
menu.

The title of the screen changes to “ADJ
READY

” and the converter is is ready for

calibration.

Pushing returns you to the previous
screen.

Push to start calibration for 0% flow
rate.

It takes several seconds to perform
calibration for 0% flow rate and the
simulated value of 0% after calibration
appears.

Pushing returns you to the setting
menu.

Note 1: To perform calibration, push and hold longer.

Note 2: To cancel the adjustment when ADJ READY is displayed, push .

The screen returns to the zero display using the simulation input.