3 100 % flow rate calibration (span calibration), 4 checking the excitation current – Toshiba LF516 User Manual

Page 106

Advertising
background image


105

6 F 8 A 0 9 3 4

9.2.2 50 % Flow Rate Calibration

Using Mag-Prover’s internal calibration circuit, 50% flow rate calibration can be performed. For
calibration procedure, see the calibration procedure for 0% flow rate. (For 50% flow rate calibration,
select "CAL 50" from the setting menu.)

9.2.3 100 % Flow Rate Calibration (Span Calibration)

Using Mag-Prover’s internal calibration circuit, 100% flow rate calibration can be performed. For
calibration procedure, see the calibration procedure for 0% flow rate. (For 100% flow rate calibration,
select "CAL 100%" from the setting menu.)

9.2.4 Checking the Excitation Current

You can monitor the exciting current value.

●Checking the exciting current value

Switch operation

Display example

Description

Select "EX DSPL" in the setting item
selection screen.

The excitation current value appears.

Pushing returns you to the setting
menu.