6 F 8 A 0 9 3 4

15. Specifications

15.1 Flowmeter Specifications

„ Overall Specifications

Measurement range in terms of flow velocity:
0–0.5 m/s to 0–10 m/s
System accuracy: when fluid conductivity is 0.1

μS/cm or more

Accuracy

Flow rate as a

percentage of range

Span 0.5 to less than 1m/s

Span 1.0 to 10 m/s

0 to less than 50%

±0.25%FS

50 to 100%

± 0.5%FS

±0.5% of rate

Note:

The accuracy above is measured under standard operating conditions at Toshiba's
calibration facility.

Fluid conductivity:

0.01 µS/cm minimum

Measurable fluid velocity:

Fluid temperature:

–10 to +120 °C

Ambient temperature: –10 to +50 °C
Dimensions and Mass: See Chapter 16, “Outline Dimensions.”

Fluid conductivity (

μS/cm)

Measurable fluid velocity

(m/s)

Meter size
40-100mm

Meter size
25mm