13.2.3 Flow rate indication is not stable.






















































Note 1: If the detector tube is not filled with

operating fluid, the flow is indefinite and
measurement is impossible. Be sure to fill
the detector tube with operating fluid
before starting measurement.

Is power supply voltage
within the specified range?

Use a power supply within the
specified range.

Are cables securely
connected?

Connect each cable securely
to the terminal board.

Is the fluid carrying
bubbles?

Design piping so that the fluid
does not carry bubbles.

Install the flowmeter away
from equipment producing
electromagnetic or
electro- static interference.

Contact your nearest Toshiba
representative.

NO

NO

YES

YES

Is

there

high-voltage

or

large

current cable or equipment
near the flowmeter?

Is the flowmeter grounded
with 100

or less ground

resistance?

Ground the flowmeter with a
copper braid or

wire(5.5

mm²

minimum)

to a

good

earth

ground

(100 or less ground

resistance).

NO

YES

YES

YES

NO

NO

Is the detector pipe filled
with fluid?

Design piping so that the
detector pipe is filled with
fluid all the time.

NO

YES

START