90

6 F 8 A 0 9 3 4

Switch operation

Display example

Description

The switches at the bottom change.
( are shown.)
At the same time, the cursor appears.
(The item indicated by the cursor is
highlighted.)

Then push .

You can continue to change the setting item.
Push to change the selection
items.

: Selected item is scrolled up.

: Selected item is scrolled down.

When the desired item is selected, push

to set the item temporarily.

When you push , a message appears
to confirm whether the setting is OK or not.
If OK, push . If you want to redo the
setting, push .

Pushing cancels the setting operation
and exits the setting screen.


8.2.23 Flow Rate High/Low Alarm and High-High/Low-Low Alarm

The high/low limit, high-high/low-low limit of the flow rate, at which an alarm is generated, can be set
as % value of the span flow rate of the set maximum range.

The high/low alarm, and high-high/low-low alarm values for flow rate can be set within the range of
−10% to 110% (percentage to Range 1) in increments of 0.1%.

●Changing the high/low alarm on/off setting

The following is an example to change the high alarm setting from OFF to ON.

Switch operation

Display example

Description

Select "H SET" from the setting item
selection menu.

The current setting (OFF in this example)
appears.

Then push .

* Pushing returns you to the menu
screen.