4) preventing an empty pipe condition – Toshiba LF516 User Manual

Page 24

Advertising
background image


23

6 F 8 A 0 9 3 4

(3) Flow Direction


Install

the

detector

in

accordance

with

the

flow

direction

arrow

on

the

detector.

See

Figure

4.7. If the actual flow runs opposite to the specified flow direction, the following
display and output appears.

For single range measurement,

LCD display:

Instantaneous flow rate ------------- indicates negative values,

Totalized flow ------------------------ no counts added.

Outut:

Current output ------------------------ 4.0mA output,

Pulse output --------------------------- No pulses

For bidirectional range measurement, the flow in opposite direction results in a positive
output value. See 10.3, “Multi-range Functions.”

Figure 4.7 Flow direction arrow on the detector

(4) Preventing an Empty Pipe Condition

Design an upright pipe run (Figure 4.8) or sufficient head pressure (Fig. 4.9) at the
downstream detector outlet if there is a possibility of the detector pipe becoming emptied.











Figure 4.8 Detector with an upright pipe run at downstream outlet










Figure 4.9 Detector with sufficient head pressure at downstream outlet

Flow direction arrow

Upright pipe run

Detector

Detector