2 external device connections and grounding – Toshiba LF516 User Manual

Page 29

Advertising
background image


28

6 F 8 A 0 9 3 4

5.2 External Device Connections and Grounding

The terminal board connections of LF516/LF546 Flowmeter are shown in Figure 5.1. Proceed with wiring
as described in Section 5.4, “Wiring Procedure.”

































Figure 5.1 External Wiring Schematic Diagram

* Use a heavy copper braid or wire (cross-sectional area 5.5 mm

2

minimum) to ground the terminal

and make it as short as possible as shown in Figure 5.1 for grounding.
Also, Avoid a common ground where earth current may flow. (An independent ground is
preferable.)

* The converter has no power switch. Install the power switch at the system side. Be sure to use a

double-pole/single-throw (both disconnection) wiring breaker.

Power supply

Wiring breaker

(double-pole/single-throw)

Output signal cable

(CVV-S)

Current output

(4-20mAdc)

or PROFIBUS

NOTE:

To avoid 2-point grounding,

ground the shield of the output

cable basically at the receiving

side.

Digital output 1
Common for DI/DO

(100 ohm or less

ground resistance)

Digital output 2

Digital input

(20 to 30 Vdc)

Power cable (CVV)

* For a 2-core cable, L1 and

L2 only.

IV wire

5.5mm

2

or more

Ground terminal

(100 ohm or less

ground resistance)

Converter unit

(100 ohm or less ground

resistance)

NOTE:

• When 3-core power cabel is used, do

not ground from the external

grounding terminal.(Perform grouding

at the receving side only).

• Only when a 2-core power cable is

used, perform grounding from the

external grounding terminal.

[Instrumentation panel: Customer]

(Receiving side)