Toshiba LF516 User Manual

Page 79

Advertising
background image


78

6 F 8 A 0 9 3 4

Switch operation

Display example

Description

Push to move the cursor to the
desired digit and push to change the
symbol or number of the digit.

Pushing shows a message to confirm
whether the setting is OK or not.
If OK, push . If you need to redo the
setting, push .

Pushing cancels the setting operation
and exits the setting screen.

Note: The manual zero adjustment value can be set within the range equivalent to ±1 m/s

(±10 % of the maximum range 10m/s). If you try to set a value out of this range, an
error message HIGH OVER SPEC or LOW OVER SPEC appears. If this happen,
redo the setting.

In addition, if you perform still water zero adjustment with water held still, the manual

zero adjustment value will be cleared to 0.0%.

8.2.16 4

20mA Alarm Output Setting

The 4

−20mA alarm output setting is the function to fix the current output to a selected fixed value if

an alarm occurs when self-diagnosis function is performed.

The 4

−20mA alarm output value can be selected from the following table.

● The 4

−20mA alarm output setting function

Selection items

The 4

−20mA alarm output value

UNDER 3mA

3.0mA or less

4mA 4.0mA

HOLD

Fixed to the present value

OVER 24mA

24.0mA or more

You can check or change the 4

−20mA alarm output value as described below.

The following is an example to change the setting from UNDER 3.0mA to 4.0mA.

Switch operation

Display example

Description

Select "ALM 4-20" from the setting item
selection menu.