10.6 Remote Zero Adjustment

On-stream zero adjustment in a zero flow rate condition can be started with an external signal.
To do this, set DI as a zero adjustment start signal. See 8.2.19, “Digital I/O”

[Signal Input Timing]











The start signal must be set to H level first, then it must go to L level after the passage of more

than 10 seconds but not more than 20 seconds, as shown above.
If the signal does not go to L level within this specified period, it will be ignored.

10.7 Remote Selection of Fixed Value Output

A user-specified 4-20 mA output and pulse output can be selected with a DI signal.
Proceed as follows to use this function:

Fixed-value setting

„

Set the fixed-value for current output and for pulse output. See 8.2.25, “Fixed-Value
Output.” Set the fixed-value output enable/disable status to “OFF.”

If the pulse output is not used, fixed-value setting for pulse output is not needed.

DI function setting

„

Set DI to use as a fixed-value output control signal. See8.2.19, “Digital Input.”

Control signal input conditions:

Control signal input level

4 –20 mA and pulse output

L level

Outputs the measured value.

H level

Outputs the fixed-value.

10 seconds

20 seconds

H level

L level

Zero adjustment

start signal

Zero adjustment start