92

6 F 8 A 0 9 3 4

Switch operation

Display example

Description

Push to move the cursor to the
digit you want to change and push
to change the number of the digit.

When you push , a message appears
to confirm whether the setting is OK or not.
If OK, push . If you want to redo the
setting, push .

Pushing cancels the setting operation
and exits the setting screen.

Note: If you try to set a value outside of the range

−10% to +110%, LOW OVER SPEC or

HIGH OVER SPEC

error appears and the value returns to the previous value. Set a

value once again.


8.2.24 Mag-Prover

-Self Diagnosis ON/OFF Setting

You can select on/off setting for Mag-Prover’s self-diagnosis function.
If the self-diagnosis function is set to OFF, no error message is displayed even if any of the errors listed
below occurs.

• ROM error

• RAM error

• System parameter error

• Excitation cable is not connected or its wiring is open

• Excitation circuit failure

• ADC circuit error

• Totalizer data is destroyed

Selection items

Description

OFF

Self-diagnosis function is turned off.

ON Self-diagnosis

function is turned on.

If this function is set to ON and an error occurs, an error message appears on the Display 2 measured
value screen. If an error occurs, the measurement items specified for Display 2 screen cannot be
displayed unless the error is removed.