1 digital i/o specifications – Toshiba LF516 User Manual

Page 108

Advertising
background image


107

6 F 8 A 0 9 3 4

10.1 Digital I/O Specifications

The specifications of the digital I/O terminals for the converter for electromagnetic flowmeter: LF546 are
as follows:

„ Digital Output 1(DO1 )

Output type:

Transistor open collector

Number of outputs:

1

Capacity:

30 V dc, 200 mA maximum

„ Digital Output 2(DO2 )

Output type:

Solidstate relay (non polarity)

Number of outputs:

1

Capacity:

150 V dc, 150 mA maximum

150 V ac(peal-to-peak), 100 mA maximum

„ Digital Input (DI )

Input signal:

20 to 30 V dc voltage signal
• High input level—20 to 30 V dc
• Low input level—2 V dc maximum

Input resistance:

Approximately 2.7 kΩ

Number of inputs:

One point

• Each I/O terminal can be used as a specified function terminal when selected.
• Terminal COM is the signal COMMON for the other three terminals (DO1, DO2 and DI).
• Each terminal is isolated from the internal circuits.

(The output terminals are not isolated from each other.)














Flowmeter

Solidstate relay

Transistor open collector
(Standard)

Voltage input (DI)

Contact output 2 (DO2)

Contact output 1 (DO1)

Resistor

Photo coupler


COM

DI

DO2

DO1

Advertising