25 fixed value output – Toshiba LF516 User Manual

Page 94

Advertising
background image


93

6 F 8 A 0 9 3 4

●Changing the Mag-Prover’s self-diagnosis function setting

The following is an example to change the Mag-Prover’s self-diagnosis setting from OFF to ON.

Switch operation

Display example

Description

Select "SELF CHK" from the setting item
selection menu.

The current setting (OFF in this example)
appears.

Then push .

* Pushing returns you to the menu
screen.

The switches at the bottom change.
( are shown.)
At the same time, the cursor appears.
(The item indicated by the cursor is
highlighted.)

Then push .

You can continue to change the setting item.
Push to change the selection
items.

: Selected item is scrolled up.

: Selected item is scrolled down.

When the desired item is selected, push

to set the item temporarily.

When you push , a message appears
to confirm whether the setting is OK or not.
If OK, push . If you want to redo the
setting, push .

Pushing cancels the setting operation
and exits the setting screen.


8.2.25 Fixed Value Output

The fixed value output function is used to output a fixed current and/or a fixed pulse output
independently of the flow rate signal. (The fixed pulse output is available only when Digital Output 1
(DO1) or Digital Output 2 (DO2) is used for pulse output function. For DO2, output can be obtained
only when fixed pulse output is 100pps or less.

The fixed-value output can be set in the ranges described below. (Current output and pulse output can
be set and output at the same time.)

• Fixed current output:

2.4 to 24 mA (can be set in increments of 0.1 mA)

• Fixed pulse output:

0 to 3000 pps (can be set in increments of 1 pps)

If fixed output is set to ON, Display 2 screen is used to indicate the fixed output in the measurement
mode.

Advertising
This manual is related to the following products: