Note 1:The meter size display loops as shown below:

25mm

40mm 50mm

80mm

100mm

1inch 1.5inch 2inch 3inch 4inch

Note 2: When the meter size is changed, range unit and count rate will be forcefully changed as

described below. If necessary, change these parameters again.

Range unit

m/s

Count rate

If the count rate goes out of the setting range
when the meter size is changed, the count rate
will be forcefully set to zero.

Note 3: The exciting frequency setting may become inappropriate for the set value when the

meter size of the detector is changed. If the exciting frequency is the value shown
below when the meter size is changed, the exciting frequency will be forcefully
changed.

Setting meter size

(mm) (inch)

Set exciting frequency

25 1

400Hz

40 ~ 80

1.5 ~ 3

200Hz

100 4

100Hz