OVER

Alarms and Errors

Status Information (con’t)

CM Status.

Stores recent error codes related to the Continuous

Monitoring mode (may be useful when calling Customer

Care).
Note: The CM Status screen shows information about

calibration and sensor errors. You can get information

about why the system did not prompt you to perform a

calibration (“Can’t Cal: xxx”) or why a calibration failed (“Cal

Failed: xxx”) or why continuous glucose readings are not

displayed (“Sen Err: xxx”). These numbers and information

help troubleshoot your system performance. The Customer

Care representative may ask you to access the CM Status

information.

Calibration BG.

Allows you to add a new blood glucose reading used for

calibration. Use this selection ONLY at the request of a

Customer Care representative.
Note: Once you use this feature to successfully perform an

additional calibration, the system discards the parameters

from the previous calibrations. It begins to calculate glucose

results based on this calibration. When using the Calibration

BG feature, always perform an additional Blood Glucose

mode test to confirm the first test.
This feature cannot be used in the first ten hours after sensor

insertion. Using this feature does not change the regular

scheduled times for calibration.

Status Information

Description

Fig. 1

Fig. 2

Display

Description

The system will notify you of any conditions that may need

your attention through messages that will be displayed on

your receiver screen. You can find out what those messages

mean and how to respond to them. For example, see Fig. 1.

The messages are arranged alphabetically in the table titled

“Respond to alarm message”.

When you use your built in FreeStyle Meter to make

measurements, you may receive error codes on the receiver.

You can find out how to respond when you get those codes.

For example, see Fig. 2.

Section 10

91

Respond to Alarms, Errors, and Problems

CM Status 08:30A

Last Reset: OK
Can’t Cal: 6
Cal Failed: BG High
Sensor Err: 207

Back Main

08:30A

Yes No

Do You Want To

Calibrate?

Glucose CM 03:00A

85

mg/dL

Acknowledge Main

Projected Low

Glucose

Glucose BG 08:30A

Er1

Home