Steps do this 5, If then – Abbott Freestyle PRT03831-xxx User Manual
Page 83
OVER
Steps
Do This
5.
Set, review, or change your
Projected Low glucose alarm.
a. Select Projected Low from the Alarms menu.
• Use the UP/DOWN Arrow buttons to highlight
Projected Low.
• Press the RIGHT Option button to select
Projected Low.
b. Review or change your settings.
• Press the LEFT Option button (Next) to move
from one field to the next.
• Use the UP/DOWN Arrow buttons to change
the setting in any field.
• Press the RIGHT Option button (Set) to accept the
settings and return to the Alarms menu.
Note: The receiver will respond with a brief sample
of the alarm you selected.
c. Decide what to do next.
• You are setting, reviewing, or changing
all alarms.
Continue to the next step.
• You are finished setting, reviewing, and
changing alarms.
Go to Step 11.
If
Then
Decide what settings you want:
• Alarm on or off.
• Sensitivity.
• Alarm type (Setting).
– Low, medium, or high beep.
– Short, medium, or long
vibration.
If you want this much notice
before reaching your Low
glucose level:
Then set the
Sensitivity to:
• Maximum detection (most
alarms, approximately 30
minutes in advance of events)
• Intermediate detection
(intermediate alarms,
approximately 20 minutes in
advance of events)
• Low detection (fewest alarms,
approximately 10 minutes in
advance of events)
high
Medium
Low
Section 8
73
Set, Review, or Change the Alarm Settings
Alarm Settings 08:30A
Low Glucose (70)
High Glucose (275)
Projected Low (On)
Projected High (Off)
Data Loss
Main Select
Alarm Settings 08:30A
Projected Low Glucose
ON
Sensitivity: Low
Setting: Medium Beep
Next Set