Steps do this 6, If then – Abbott Freestyle PRT03831-xxx User Manual

Page 84

Advertising
background image

Steps

Do This

6.

Set, review, or change your

Projected High glucose alarm.

a. Select Projected High from the Alarms menu.

• Use the UP/DOWN Arrow buttons to highlight

Projected High.

• Press the RIGHT Option button to select

Projected High.

b. Review or change your settings.

• Press the LEFT Option button (Next) to move

from one field to the next.

• Use the UP/DOWN Arrow buttons to change

the setting in any field.

• Press the RIGHT Option button (Set) to accept the

settings and return to the Alarms menu.

Note: The receiver will respond with a brief sample

of the alarm you selected.

c. Decide what to do next.

Decide what settings you want:
• Alarm on or off.
• Sensitivity.
• Alarm type (Setting).

– Low, medium, or high beep.
– Short, medium, or long

vibration.

• You are setting, reviewing, or changing

all alarms.

Continue to the next step.

• You are finished setting, reviewing, and

changing alarms.

Go to Step 11.

If

Then

If you want this much notice

before reaching your High

glucose level:

Then set the

Sensitivity to:

• Maximum detection (most

alarms, approximately 30

minutes in advance of events)

• Intermediate detection

(intermediate alarms,

approximately 20 minutes in

advance of events)

•Low detection (fewest alarms,

approximately 10 minutes in

advance of events)

high

Medium

Low

Customer Care: 1-866-597-5520

74

Section 8

Set, Review, or Change the Alarm Settings

Alarm Settings 08:30A
Low Glucose (70)
High Glucose (275)
Projected Low (On)

Projected High (Off)

Data Loss

Main Select

Alarm Settings 08:30A
Projected High Glucose

ON

Sensitivity: Low

Setting: Medium Beep
Next Set

Advertising