Steps
Do This
3.
Set, review, or change the
Low Glucose alarm.
a. Select Low Glucose from the Alarms menu.
• Use the UP/DOWN Arrow buttons to highlight
Low Glucose.
• Press the RIGHT Option button to select Low
Glucose.
b. Review or change your settings.
• Press the LEFT Option button (Next) to move from
one field to the next.
• Use the UP/DOWN Arrow buttons to change
the setting in any field.
• Press the RIGHT Option button (Set) to accept the
settings and return to the Alarms menu.
Note: The receiver will respond with a brief sample
of the alarm you selected.
c. Decide what to do next.
• You are setting, reviewing, or changing
all alarms.
Continue to the next step.
• You are finished setting, reviewing, and
changing alarms.
Go to Step 11.
If
Then
Decide what settings you want:
• Alarm on or off.
• Your threshold value.
Note: The alarm can only be set
between 60 and 139 mg/dL
(3.3 mmol/L and 7.7 mmol/L).
• Alarm type (Setting).
– Low, medium, or high beep.
– Short, medium, or long
vibration.
Caution: The Low Glucose alarm cannot be set below
60 mg/dL (3.3 mmol/L). Therefore, it is not intended to
notify you of severe hypoglycemia.
Section 8
71
Set, Review, or Change the Alarm Settings
Alarm Settings 08:30A
Mute Alarms
Low Glucose (70)
High Glucose (275)
Projected Low (On)
Projected High (Off)
Main Select