Steps do this 3, If then – Abbott Freestyle PRT03831-xxx User Manual

Page 81

Advertising
background image

OVER

Steps

Do This

3.

Set, review, or change the

Low Glucose alarm.

a. Select Low Glucose from the Alarms menu.

• Use the UP/DOWN Arrow buttons to highlight

Low Glucose.

• Press the RIGHT Option button to select Low

Glucose.

b. Review or change your settings.

• Press the LEFT Option button (Next) to move from

one field to the next.

• Use the UP/DOWN Arrow buttons to change

the setting in any field.

• Press the RIGHT Option button (Set) to accept the

settings and return to the Alarms menu.

Note: The receiver will respond with a brief sample

of the alarm you selected.

c. Decide what to do next.

• You are setting, reviewing, or changing

all alarms.

Continue to the next step.

• You are finished setting, reviewing, and

changing alarms.

Go to Step 11.

If

Then

Decide what settings you want:
• Alarm on or off.
• Your threshold value.
Note: The alarm can only be set

between 60 and 139 mg/dL

(3.3 mmol/L and 7.7 mmol/L).
• Alarm type (Setting).

– Low, medium, or high beep.
– Short, medium, or long

vibration.

Caution: The Low Glucose alarm cannot be set below
60 mg/dL (3.3 mmol/L). Therefore, it is not intended to
notify you of severe hypoglycemia.

Section 8

71

Set, Review, or Change the Alarm Settings

Alarm Settings 08:30A
Mute Alarms

Low Glucose (70)

High Glucose (275)
Projected Low (On)
Projected High (Off)

Main Select