18 Appendix B: Specifications

(con’t)

Determination of Alarm Performance

The threshold alarm is characterized below. The projected alarm performance has not been established.
The performance of low and high glucose alarms was assessed in an in-clinic study using 58 subjects with

type 1 diabetes wearing one FreeStyle Navigator sensor on the arm and one sensor on the abdomen. FreeStyle

Navigator continuous data were masked from the subjects and investigators and the alarms were not turned

on. During 50 hours the subjects’ venous glucose was tested with a YSI 2300 Stat Plus glucose analyzer at 15

minute intervals. Arm and abdomen data were pooled in the alarm analysis. Alarm performance was evaluated

in a retrospective analysis of the study data. As alarm performance was developed retrospectively, your results

may vary from those reported below.
Definitions:

Hypoglycemic event – two or more successive YSI measurements below the alarm threshold or one YSI

measurement 6 mg/dL below the alarm threshold.
Hyperglycemic event – two or more successive YSI measurements above the alarm threshold or one YSI

measurement 6% above the alarm threshold.
True Threshold Alarm – a threshold alarm that occurred ± 30 minutes from the start of a hypoglycemic or

hyperglycemic event
True Alarm Rate – the percentage of time the glucose level was beyond the threshold and an alarm was

activated

Events Detected by True Threshold Alarms

Total Events

X 100

Missed Alarm Rate – the percentage of time the glucose level was beyond the threshold and an alarm was not

activated

Events Not Detected by True Threshold Alarms

Total Events

X 100

False Threshold Alarm – a threshold alarm that occurred when a YSI measurement within ± 30 minutes was

not beyond the threshold setting

False Alarm Rate – the percentage of time an alarm occurred when glucose level was not beyond the threshold

setting

False Threshold Alarms
Total Threshold Alarms

X 100