Steps do this 1, If then – Abbott Freestyle PRT03831-xxx User Manual
Page 66
Steps
Do This
1.
Insert a FreeStyle Test Strip into
the receiver’s test strip port.
a. Insert the end with the dark rectangle.
b. Gently push until the test strip stops.
Result: The receiver will display the most recently used
test strip code number.
2.
Ensure that the code number
on the screen matches the code
number on the test strip vial.
Compare the code number on the display with the code
number on the vial of test strips.
Result: The receiver will prompt you to “Apply Sample.”
• The numbers match. a. Press the RIGHT Option button (Set).
b. Go to Step 3.
• The numbers do
NOT match.
a. Use the UP/DOWN Arrow buttons to change the
code to match.
b. Press the RIGHT Option button (Set).
c. Go to Step 3.
• The screen changes
to the Apply Sample
screen.
If the numbers do NOT match:
a. Press the LEFT Option button (Back).
b. Repeat Step 2.
If the numbers match, go to Step 3.
If
Then
Caution:
The code numbers MUST match to ensure
accurate test results.
Customer Care: 1-866-597-5520
56
Section 7
Calibrate Your System/Test Blood Glucose Manually
Strip Code
08:30A
Cancel Set