The receiver will not turn on – Abbott Freestyle PRT03831-xxx User Manual

Page 112

Advertising
background image

a. Was the test strip inserted correctly (the “FreeStyle” side

was up as you inserted the end with the dark rectangle)?
If YES, go to Step c.
If NO, repeat the test with a new strip taking care that

the “FreeStyle” side is up as you insert the end with the

dark rectangle until it stops.

b. Did this solve your problem?

If YES, stop.
If NO, continue with Step c.

c. Are the batteries installed correctly?

If YES, go to Step d.
If NO, Install the batteries correctly and reset time

and date.

d. The batteries may be dead, replace existing batteries

and reset time and date.

e. Repeat the test with a new test strip.

If this did not resolve your problem call Customer Care.
Note: Whenever batteries are removed and replaced or

re-installed in the receiver, you will have to reconnect

the transmitter and receiver and insert a new sensor.

The receiver will not turn on.

Note: Ensure that you apply your sample when the
Apply Sample message appears on the display.
a. Repeat the test using a LARGER sample on a new

test strip.

b. If this did not resolve your problem call Customer Care.

The Blood Glucose test won’t start after you apply a sample.

Customer Care: 1-866-597-5520