Steps do this 7 – Abbott Freestyle PRT03831-xxx User Manual

Page 41

Advertising
background image

Steps

Do This

7.

Evaluate the test results.

Compare the test results displayed on the screen to the
range printed on the vial of test strips.

8.

Check your work.

Result: Control solution test results displayed on the
receiver screen.
You have done this procedure correctly when:

o

The results are within the normal control solution range.

o

The used test strips have been discarded properly.

o

The control solution tests are marked as control

solution tests.

END OF Section 4

If the result

And

Then

• Falls WITHIN the

range.

Go to Step 8.

• Falls OUTSIDE of

the range.

You’ve repeated

the test.

Call Customer Care.

You’ve NOT

repeated the test.

Repeat the test from Step 1 with a

NEW test strip.

• Is an error message.

Go to Section 10 on

troubleshooting.

77 – 115 mg/dL

Section 4

31

Perform a Control Solution Test

Glucose BG 08:30A

Yes No

106

mg/dL

Control Solution?

Advertising