Steps do this 7 – Abbott Freestyle PRT03831-xxx User Manual
Page 41
Steps
Do This
7.
Evaluate the test results.
Compare the test results displayed on the screen to the
range printed on the vial of test strips.
8.
Check your work.
Result: Control solution test results displayed on the
receiver screen.
You have done this procedure correctly when:
o
The results are within the normal control solution range.
o
The used test strips have been discarded properly.
o
The control solution tests are marked as control
solution tests.
END OF Section 4
If the result
And
Then
• Falls WITHIN the
range.
➜
Go to Step 8.
• Falls OUTSIDE of
the range.
You’ve repeated
the test.
Call Customer Care.
You’ve NOT
repeated the test.
Repeat the test from Step 1 with a
NEW test strip.
• Is an error message.
➜
Go to Section 10 on
troubleshooting.
77 – 115 mg/dL
Section 4
31
Perform a Control Solution Test
Glucose BG 08:30A
Yes No
106
mg/dL
Control Solution?