Follow the guidelines below when selecting your alarm settings for detection of high and low glucose

conditions. When the detection rate is maximum, you will get the greatest number of alarms. When

the detection rate is minimum, you will get the fewest number of alarms.

About Glucose Alarms (con’t)

Caution: Do NOT rely solely on the projected glucose alarms for detection of low or high glucose conditions.

Always use both the low glucose and high glucose threshold alarms AND the projected low and projected

high glucose alarms for maximum notification of low or high glucose conditions.

Threshold

Projected

Detection

Alarm

Alarm

Rate

ON

ON

Maximum

High Sensitivity (~30 min)

ON

ON

Intermediate

Low Sensitivity (~10 min) or

Medium Sensitivity (~20 min)

ON

OFF

Minimum

OFF

OFF

None