18 appendix b: specifications, Con’t) – Abbott Freestyle PRT03831-xxx User Manual

Page 167

Advertising
background image

OVER

18 Appendix B: Specifications

(con’t)

Performance Over the Duration of Wear
This section presents the performance data in a variety of ways to demonstrate the performance of

the system over time during the entire sensor wear period. The sensor is worn on the body for up

to 5 days, during which time the system has to be calibrated 4 times. Typically, the system has to be

calibrated at 10, 12, 24 and 72 hours after sensor insertion. Data presented in this section demonstrates

how the sensor performs as a function of time.
Sensor Stability

Tables 6a-6b show there is little difference in accuracy over the five days of sensor wear according to

the Clarke Error Grid Analysis (Table 6a) and the Continuous Glucose-Error Grid Analysis (Table 6b).
Table 6a. Clarke Error Grid Analysis by day of wear

Zone

Day 1

Day 2

Day 3

Day 4

Day 5

%

%

%

%

%

A

82.5

82.4

79.4

84.0

80.9

B

16.4

16.6

18.3

14.2

16.9

C

0.2

0.1

0.0

0.0

0.0

D

0.9

0.9

2.2

1.8

2.1

E

0.0

0.1

0.0

0.0

0.0

Overall Mean Absolute

Relative Difference (%)

MARD0

12.6

12.3

14.1

11.9

13.0

Table 6b. Continuous Error Grid Analysis by day of wear

Zone

Day 1

Day 2

Day 3

Day 4

Day 5

%

%

%

%

%

Accurate

Readings

97.9

97.8

97.1

97.4

97.3

Benign

Errors

1.0

1.2

0.7

0.8

0.8

Erroneous

Readings

1.1

1.0

2.2

1.8

1.9

Table 6c groups the difference between glucose results from the FreeStyle Navigator system and the

YSI into different blocks (within 20%, within 30% and within 40% from the YSI). It shows there is little

change in the difference from YSI over time, thus demonstrating sensor stability.
Table 6c.

Difference from the YSI at different times after sensor insertion

Time After Insertion

(hours)

Percent Readings

Within 20% of the

YSI*

Percent Readings

Within 30% of the

YSI*

Percent Readings

Within 40% of the

YSI*

10-12

86

92

96

12-24

82

93

97

24-72

81

92

97

72-122

82

93

97

*The absolute difference from the YSI reading is measured in mg/dL if the YSI reading is at or below

75 mg/dL.

Section 18

157

Appendix B: Specifications