Steps

Do This

5.

Apply control solution.

Important: Do NOT apply control solution to both edges of
the test strip.

Notes:
• Gently touch ONLY ONE EDGE of the test strip to the

control solution next to one of the dark half-circles.

• If progress tones are turned on, the receiver will beep to

indicate that enough control solution was applied.

Result: A stopwatch

will appear on the display

screen while the receiver measures the control solution.

• The receiver will beep twice when the test results appear.

6.

Confirm that you used a

control solution.

Press the LEFT Option button (Yes) to alert the receiver
that you used a control solution for this test.

Note: If you do NOT select yes, the system will record the

result as a blood glucose result instead of a control solution

result, which can result in an incorrect statistical report.

Go to the next step.

Customer Care: 1-866-597-5520

30

Section 4

Perform a Control Solution Test

Apply Sample

08:30A

Back Cancel

Glucose BG

08:30A

Yes No

Control Solution?

106

mg/dL