OVER

Steps

Do This

4.

Prepare for lancing. (con’t)

g. Select the lancing device depth setting.

• The FreeStyle® Lancing Device offers four different depth

settings. Move the dial to the desired setting as shown

in the depth indicator window. Level 1 is the shallowest

depth; Level 4 is the deepest. To lance parts of the body

other than the finger, we recommend that you start at

Level (2) and use the clear cap.

To lance your finger, we recommend that you set the

depth setting to its shallowest depth (1) and use the

grey cap.

h. Pull the dark grey cocking handle out until it clicks.

Note: You may have already cocked the device in Step 4d.

i. Prepare the site:

• A finger.

Set the depth setting to its shallowest depth (1).

• Your forearm.

Start at Level (2) or higher if your experience with

setting 2 does not produce a usable sample.

I

f you are lancing

Then

• A finger.

Warm your hand, lower it to waist level, and

gently massage the finger.

• Your forearm.

Rub the test site vigorously until you feel it getting

warm (three to five seconds).

If you are using

Then to stimulate blood flow

Section 7

59

Calibrate Your System/Test Blood Glucose Manually