Settings/setup

System 2020 XT USB

5 - 22

ba76019e01

12/2012

5.7.2

Erasing a Sensor-sensor link

Note
The Sensor-sensor link function is automatically deactivated if the
measured parameter of the linked sensor is changed (e.g. by switching
the display of the TetraCon

®

700 IQ conductivity sensor from salinity to

conductivity).

1

Using <S>, open the Settings menu.

2

Using <

><> and <OK>, select and confirm the menu item,

System settings -> Sensor-sensor link.
The Sensor-sensor link overview opens.
The displayed list includes all sensors with which the
correction of an influence quantity is possible (e.g. FDO

®

700

IQ).

3

Using <

><>, highlight a sensor and confirm with <OK>.

The displayed list includes all linked sensors.

4

Using <

><>, highlight a sensor and confirm with <OK>.

The displayed list includes all sensors that measure an
influence quantity with the correct unit
(e.g. TetraCon

®

700 IQ, which measures the salinity in mg/L),

and the menu item, Erase link.

5

Using <

><>, highlight the menu item, Erase link, and

confirm with <OK>.
The sensor-sensor link is deleted.

6

Leave the system settings with <M>.
In the measured value display, the measured value is
corrected by the manually entered value.