YSI IQ SensorNet 2020 XT Terminal User Manual

Page 114

Advertising
background image

Operation

System 2020 XT USB

4 - 36

ba76019e01

12/2012

The restoration of a configuration onto the main controller is carried out
manually from the USB memory device. A backup of the configuration
onto USB memory device can be carried out in one of the following
ways:

 Regular manual backup to USB memory device

 Regular automatic backup to MIQ/TC 2020 XT (configured as

terminal)

If no current manual backup is present, the automatic backup of the
MIQ/TC 2020 XT can be used.
To transfer the configuration automatically backed up on the MIQ/
TC 2020 XT to a USB memory, operate the MIQ/TC 2020 XT as a
controller or backup controller. In this status the configuration data of
the backup controller can be backed up onto a USB memory as a
manual backup.

Manually restore after

MIQ/TC 2020 XT

controller operation

To transfer configuration data onto the controller, proceed as follows:

1

Using <S>, open the Settings menu.

2

Using <

>, select the menu item, Data transfer to USB

memory and confirm with <OK>.

3

Using <

>, select the menu item, Retransfer

configuration and confirm with <OK>. The process is started.

4

Wait for the restore process to be completed.

5

Confirm the concluding message with <OK>.