YSI IQ SensorNet 2020 XT Terminal User Manual
Page 141

System 2020 XT USB
Settings/setup
5 - 21
ba76019e01
12/2012
Behavior if there is no
measured value for the
influence quantity
5
Using <
><>, highlight a sensor that can be linked, and
confirm with <OK>.
The sensors are linked.
The influence quantity to be linked and the linked sensor
are displayed.
6
Leave the system settings with <M>.
In the measured value display, the corrected measured value
is marked by an asterisk (*).
Cause
Behavior
– Sensor failure
– The manually entered value for the
influence quantity is automatically used for
correction.
An exclamation mark (!) indicates the
interrupted link.
– The sensor-sensor link is reactivated as
soon as the measured value for the
influence quantity is available again.
– Maintenance
condition active
– The value last measured for the influence
quantity is automatically used for
correction.
– The current measured value is used for
correction as soon as the measured value
for the influence quantity is available again.