YSI IQ SensorNet 2020 XT Terminal User Manual

Page 141

Advertising
background image

System 2020 XT USB

Settings/setup

5 - 21

ba76019e01

12/2012

Behavior if there is no

measured value for the

influence quantity

5

Using <

><>, highlight a sensor that can be linked, and

confirm with <OK>.
The sensors are linked.
The influence quantity to be linked and the linked sensor
are displayed.

6

Leave the system settings with <M>.
In the measured value display, the corrected measured value
is marked by an asterisk (*).

Cause

Behavior

– Sensor failure

– The manually entered value for the

influence quantity is automatically used for
correction.
An exclamation mark (!) indicates the
interrupted link.

– The sensor-sensor link is reactivated as

soon as the measured value for the
influence quantity is available again.

– Maintenance

condition active

– The value last measured for the influence

quantity is automatically used for
correction.

– The current measured value is used for

correction as soon as the measured value
for the influence quantity is available again.