11.2.2 Info messages
EI8141
Connection to the component instable
* Check installation and cable lengths,
Follow installation instructions
* Set SN terminator switch acc. to operating manual
* Check environmental effects
* Component defective, contact service
EI9141
Power failure occurred
* Check date and time and, if necessary, adjust them
Message code
Message text
IA2141
Automatic air pressure measurement component identified
The measured values of the air pressure compensation component
are used
II1141
Language not available,
Default language German
* Contact service
II2141
New IQ Sensor Net component identified
II3141
New IQ Sensor Net component recorded
* See component lists
II4141
IQ Sensor Net component recorded as replacement component
* See component lists
II5141
Link sensor - output has been erased
* if necessary, link sensor again
II9141
Date and time have been set
Message code