System 2020 XT USB

Settings/setup

5 - 31

ba76019e01

12/2012

Alarm links setting table

Menu item

Selection/Values

Explanations

Measured variable

Main variable
Adjoining variable

Main variable designates the actual
measured parameter of the sensor (e.g.
pH, oxygen, etc.).
Adjoining variable designates an
additional measured parameter (e.g.
temperature).

Limit value

Lower limit
Upper limit

Type of the alarm event.

Lower limit: An alarm is triggered if the
defined limit value is undercut.

Upper limit: An alarm is triggered if the
defined limit value is exceeded.

Lower limit /

Upper limit

within the measuring
range (sensor-
dependent)

Limit value for the alarm event

Hysteresis

0 - 10 % of the
measuring range

Hysteresis for the limit value

Designation

(max. 20 characters)

User-defined designation for easier
identification in the alarm message
(display and SMS).

Relay output

Dxx / .../Ry

...

...

...

No relay output

Opens a list with all relay outputs where
the Alarm contact is set.

Dxx: number of the output module
.../Ry: relay output channel

Here you can select a relay output. When
the alarm event occurs, it carries out the
specified action (Open or Close). Details
see section 5.10.3 A

LARM

OUTPUT

TO

DISPLAY

.

Accept

The settings are taken over in the setting
table by pressing <OK>. The display
switches to the next higher level.

Cancel

The display switches to the next higher
level without storing the new settings.