YSI IQ SensorNet 2020 XT Terminal User Manual

Page 165

Advertising
background image

System 2020 XT USB

What to do if ...

7 - 5

ba76019e01

12/2012

Malfunctions despite

adequate supply voltage

on both +U V wires

Component is not

registered on the

system despite

sufficient supply voltage

on both +U V wires

(with no error message)

Cause

Remedy

– Signal transmission faulty

– For terminal components,

check/correct their seating on
the MIQ module.

– Check the log book for error

messages (for details on the
log book, see section 4.5).

– Check/correct the position of

the SN terminator switches (for
the correct settings, see
section 3.9.1).

– Check the total line length

within the IQ S

ENSOR

N

ET

. For

a total line length of more than
1000 m, an MIQ/JBR signal
amplifier module is required.

– Component is defective

– Check the log book for error

messages (for details on the
log book, see section 4.5).

– If no log book entry exists, send

the component back.

Cause

Remedy

IQ sensors

– IQ sensor connection cable is

not correctly connected or
defective

– Check IQ sensor cable

connection on the terminal
strip.

– Check IQ sensor cable and, if

necessary, replace it. Another
IQ sensor can be used for the
test.

– IQ sensor is defective

– Test IQ sensor at another

measurement location.

– If IQ sensor still does not

function, contact YSI.

Other components

– Component is defective

– If possible, test component at

another measurement location.

– If component still does not

function, contact YSI.