YSI IQ SensorNet 2020 XT Terminal User Manual

Page 167

Advertising
background image

System 2020 XT USB

What to do if ...

7 - 7

ba76019e01

12/2012

"no sensor" in the

measured value display

though an IQ sensor is

connected

The selected system

language was not

activated for all

components

Cause

Remedy

– Local display has been set up

and the IQ sensor is not included
in the local display

– Add the sensor to the local

display (see section 5.9)

– Switch to the All sensors

display with Display local
values or all values
(see
section 4.4.5)

Cause

Remedy

– A system language was selected

that is not available in one
component (sensor, controller,
terminal, output module).
The standard language, English,
was activated instead of the
selected system language.

– Contact YSI,

a software update is
required for the respective
components