Technical data

System 2020 XT USB

8 - 2

ba76019e01

12/2012

Warning

A circuit (except for the power supply connections), that is
connected to an IQ S

ENSOR

N

ET

component, must not feed any

voltages or currents that are not allowed. It has to be made sure
that the circuit at any time meets all requirements of a Limited
circuit
or Limited Power as well as of SELV (Safety Extra Low
Voltage). These include the following limiting value
specifications:
AC voltage: max. 30 V effective / 42.4 V peak
DC voltage: max. 60 V
Current limit: max. 8 A
Power output limitation: max. 150 VA

Internal voltage supply

Relationship between the supply voltage U, the LED status on the
MIQ module and the log book entries (Fig. 8-1):

Fig. 8-1

Relationship of supply voltage - LED states - log book

LED state

Log book

LED yellow

No entry (= OK)

LED red

1st error message

Both LEDs off

2nd error message (or switch-off)

U [V]

24

18

16,4
14,4
14,2

LED yellow

LED yellow

LEDs off

LED red

LED red

LED red

t