1 setting the recording interval ( dt – YSI IQ SensorNet 2020 XT Terminal User Manual
Page 157

System 2020 XT USB
Settings/setup
5 - 37
ba76019e01
12/2012
5.12.1 Setting the recording interval (dt) and recording duration
(Dur.)
Fig. 5-21 Measured value logging
1
Using <S>, open the Settings menu.
2
Using <
> and <OK>, select and confirm the menu
item, Measured value logging -> .
The Measured value logging display opens.
It contains a list of all sensors and inactive sensors.
3
Press <
> and <OK> to select and confirm the dt
column.
4
Press <
> and <OK> to select and confirm a sensor.
5
Determine the recording interval with <
> and <OK>.
When the recording interval is changed, the recording duration
changes at the same time (Dur.), because the recording
duration results from the recording interval due to a fixed
number of memory blocks.
6
Switch to the selection of columns with <ESC>.
7
Press <
> and <OK> to select and confirm the Dur.
column.
8
Press <
> and <OK> to select and confirm a sensor.
9
Select the recording duration with <
> and <OK>.
The recording duration is always increased or decreased in the
same steps.