3 alarm output to display, 3 alarm output to display -32, See section 5.10.3 a – YSI IQ SensorNet 2020 XT Terminal User Manual

Page 152: Larm, Output, Display

Advertising
background image

Settings/setup

System 2020 XT USB

5 - 32

ba76019e01

12/2012

5.10.3 Alarm output to display

When an alarm event occurs, a window with a text message appears.

Fig. 5-18 Example of an alarm message on the display

The alarm message contains the following information:

Operating instructions

If there are several alarm messages indicated on the display, you can
scroll through the messages with <

>. This can be recognized

from the page number in the right lower corner. The latest message is
always on the first position.

Pressing <M> hides the alarm messages and switches to the
measured value display. After one minute the alarm messages appear
again if what caused them is still present.

 Line 1:

Alarm no. Axx and time and date of the alarm event

 Line 2:

User-defined designation

 Line 3:

Sensor number and model name of the sensor that
triggered the alarm event

 Line 4:

Series number and name of the sensor that triggered the
alarm event

 Line 5:

Description of the event with specification of the limiting
value:
"<" = undercut
">" = exceeded

Advertising