1 setting the recording interval ( dt – YSI IQ SensorNet 2020 XT Terminal User Manual

Page 157

Advertising
background image

System 2020 XT USB

Settings/setup

5 - 37

ba76019e01

12/2012

5.12.1 Setting the recording interval (dt) and recording duration

(Dur.)

Fig. 5-21 Measured value logging

1

Using <S>, open the Settings menu.

2

Using <

> and <OK>, select and confirm the menu

item, Measured value logging -> .
The Measured value logging display opens.
It contains a list of all sensors and inactive sensors.

3

Press <

> and <OK> to select and confirm the dt

column.

4

Press <

> and <OK> to select and confirm a sensor.

5

Determine the recording interval with <

> and <OK>.

When the recording interval is changed, the recording duration
changes at the same time (Dur.), because the recording
duration results from the recording interval due to a fixed
number of memory blocks.

6

Switch to the selection of columns with <ESC>.

7

Press <

> and <OK> to select and confirm the Dur.

column.

8

Press <

> and <OK> to select and confirm a sensor.

9

Select the recording duration with <

> and <OK>.

The recording duration is always increased or decreased in the
same steps.