What to do if ...

System 2020 XT USB

7 - 4

ba76019e01

12/2012

7.2.3

Tips for clearing errors in the voltage supply

Note
Admissible values for the internal voltage supply (+U) are given in
section 8.1.

Supply voltage not

present or in the

warning or error range

Cause

Remedy

– Number of power MIQ power

supply modules is not
sufficient

– Check power rating. If

necessary, install another
MIQ power supply module (for
details of how to do this, see
section 3.4.2 D

RAWING

UP

THE

POWER

RATING

).

– Cable section is too long

– Check length of the cable

section. If necessary, install
another MIQ power supply
module (for details of how to do
this, see section 3.4.3 E

FFECT

OF

THE

CABLE

LENGTH

).

– Electrical connection

between MIQ power supply
module and component is
defective
(distributed mounting)

– Check cable connection step-

by-step starting with the
MIQ power supply module, and
replace defective cable
sections.

– Check contacts on the

MIQ modules (stack mounting).
Clean any dirty contacts.
Carefully bend back contact
springs that have been pressed
flat or bent (pay attention to
sufficient spring tension).

– Tighten all screws on the

terminals.

Advertising