3 other errors, Other errors -6 – YSI IQ SensorNet 2020 XT Terminal User Manual

Page 166

Advertising
background image

What to do if ...

System 2020 XT USB

7 - 6

ba76019e01

12/2012

7.3

Other errors

Status LED

of the MIQ/TC 2020 XT

flashes

The system does no

longer react on entries

"Error" in the measured

value display

Local settings cannot be

displayed

Cause

Remedy

– A fault is present

Fault remedy according to the
logbook:
– More detailed information on

the current fault and its
remedy is given in the
logbook (see section 4.5.2)

Cause

Remedy

– System error

Reset the system:
– Remove the Terminal/

Controller and dock it again
after 10 s

or:

– Switch off the power supply

and switch it on again after
10 s

Cause

Remedy

– Communication with the

IQ Sensor interrupted

– Check the cable connection

– Error in the IQ sensor

– Disconnect the IQ sensor

and reconnect it after 10 s

Cause

Remedy

– The contacts of two

MIQ modules are not correctly
connected to each other

– Clean the contacts