4 behavior of the iq sensor net in case of error, Behavior of the iq – YSI IQ SensorNet 2020 XT Terminal User Manual

Page 24

Advertising
background image

Overview

System 2020 XT USB

1 - 18

ba76019e01

12/2012

1.4

Behavior of the IQ S

ENSOR

N

ET

in case of error

1.4.1

Behavior of the IQ S

ENSOR

N

ET

in case of power failure

 The system configuration (Settings of sensors and diff. sensors and

Settings of outputs and links) is stored permanently

 All stored measured data remain permanently available

 Linked current outputs and relay outputs switch to the non-active

condition.
Current output: 0 mA current
Relay output: Relays open
For details, refer to the component operating manual of the current
or relay output module.

 When the power supply is available again, the system is

automatically restarted, starts working again and switches to the
measured value display

 During the restart, an entry informing of the power failure is made in

the log book.

1.4.2

Behavior of the IQ S

ENSOR

N

ET

if a components fails

 If the operating voltage is too low, the LEDs on the MIQ modules

extinguish.

 The blue status LED of the MIQ/TC 2020 XT flashes

 If active components (e.g. sensors or output modules) cannot be

contacted, an entry is made in the log book. The error symbol
flashes on the display.

 The Edit list of sensors or Edit list of outputs display shows a

question mark "?" for the component.

Component

Behavior

IQ sensor

 The configuration for the IQ sensor (Settings of

sensors and diff. sensors) remains stored.

 Current measured values are not available.
 The measured value display shows ERROR.

MIQ modules

 If an active module fails, an entry is made in the

log book. The error symbol flashes on the
display.

 The configuration for these modules is retained

(e.g. links with output modules).