3.8

Connecting the voltage supply

Note
How to connect the voltage supply is comprehensively described in the
operating manual of the MIQ power supply module.

3.9

Commissioning

3.9.1

Topology and terminator switch

For failure-free operation, the terminator switches (terminating
resistors) must always be set to ON on two MIQ modules. On which
modules this has to be done can be taken from the topology of the
IQ S

ENSOR

N

ET

system:

Fig. 3-26 Correct setting of the terminator switches

Main line

Longest cable section with terminator switches at
its ends.
The terminator switches on both ends must be
set to ON.
If all MIQ modules are at one location, i. e. in a
module stack, the length of the main line is zero
(star-shaped wiring).

Branch line

All lines that branch from the main line.
Branch lines can be branched further.

x

x

s

s

s

s

s

s

s

Trunk line (”longest cable section”)

Star-shaped wiring

Branch line

Branch line 1

Branch line 2

Branch line 3
(with subbranches)

s

x

MIQ/... with terminator switch OFF

MIQ/... with terminator switch ON

Sensor

x x