3 connecting the liquid containers – YSI P 700 IQ Analyzer User Manual

Page 42

Advertising
background image

42

ba76133e01

09/2013

P 700 IQ

8

Tighten the cable glands on the power supply box.

9

Mount the lid of the power supply box.

3.3.5.3 Connecting the liquid containers

1

Place the chemicals tray on the bottom of the analyzer.

2

Mount the lids of the lid set on the luer connectors. The lids are
color coded.

3

Put on your personal protective equipment (PPE) and chemical
resistant gloves (see section 2.3 Personal protective equipment
(PPE),  19)
.

4

Open the lid of the liquid container to be connected.

5

Place the liquid container in the chemicals tray and screw the rel-
evant preassembled lid on the container. The lids with the color
coded connectors are included in the scope of delivery. The color
coding of the container must correspond to the color coding of the
lid (see Fig. 3-19,  42)!

Fig. 3-19: Chemicals container (connected)

1 Color coding of the preassembled lid
2 Color coding of the container
3 Chemicals tray

With installation in the open and ambient temperatures
below 0 °C, make sure that the liquids do not freeze during
the commissioning. If necessary, switch on the power
supply box beforehand and let the housing warm up with
the doors closed (see section 3.4 Initial commissioning,

44).