4 changing the pump tube of the peristaltic pump, See section 5.4.4,  83), Changing the pump tube of the peristaltic pump – YSI P 700 IQ Analyzer User Manual

Page 83: See section 5.4.4,  83

Advertising
background image

ba76133e01

09/2013

83

P 700 IQ

10

Reconnect the plug with the P700IQ-PO4.
or
Switch the analyzer on again with the power switch.

Restarting the

measuring operation

11

Carry out the steps 13…20 in section 5.3.4 Maintenance and
cleaning work on the open P 700 IQ,  73.

5.4.4 Changing the pump tube of the peristaltic

pump

How it functions

The liquids are pumped through the P 700 IQ with the aid of a peristaltic
pump.

A peristaltic pump is a displacement pump for the pumping of liquids.
The liquid is in a flexible tube that is fitted in a circular pump housing. A
rotor with several rolls presses down some parts of the flexible tube.
When the rotor is turning, the rolls forward the liquid through the tube.

• The pump tube has a working life of 1500 hours for water. The

chemicals used may, however, reduce the working life.

• The deterioration of the tube quality is a slow process. When the

working life of the pump tube is near its end, the P 700 IQ may have
to be calibrated more often.

• A peristaltic pump can be blocked when it was not used for a longer

period of time, because the parts of the flexible tube that are pressed
down stick together. After a shutdown of several weeks the blockage
is normally cleared after some rotations. After a shutdown of several
months the flexible tube has to be removed and the blockage
cleared by hand.

Maintenance interval

12 months

Parts required

• Pump tube PT (see chapter 7. Technical data,  99)

• Mounting aid

Preparations

1

Carry out the steps 1…11 in section 5.3.4 Maintenance and
cleaning work on the open P 700 IQ,  73.

2

Put on protective clothing (lab coat, protective goggles, chemical
resistant protective gloves).

Advertising