Expanded programs – Hach-Lange DR 6000 User Manual User Manual
Page 103

103
Expanded programs
λ/ Absorbance Formula:
1. Press Absorbance Formula.
2. The formula selected in the top key determines the number
of wavelength and coefficient keys that will appear below.
3. Press the key to change the calculation formula.
4. Select a formula from the displayed list.
5. Press OK.
When a new formula is selected, the number of variables
below changes to match.
Table 13 Multi wavelength setup options
Options
Description
More
For further Options
Recall measurement
data symbol
Recall saved measurement data, wavelength scans or time courses, refer to
% Trans/Abs
Switches from % transmittance to concentration or absorbance in reading mode.
λ
Wavelength
To input the wavelength setting. Use the alphanumeric keypad to enter the reading wavelengths.
A wavelength in the range of 190–1100 nm can be entered.
Timer icon
This functions as a stopwatch. It helps to make sure that the steps of an analysis are correctly
timed (e. g. reaction times, wait times, etc., can be exactly specified). When the specified time
has elapsed, a sound is emitted. The use of the timer has no influence on the reading program.
Concentration Factor
Multiplication factor for converting absorbance values into concentration values.
Concentration
Resolution
To select the number of decimal places.
Absorbance Formula
Calculation basis for evaluating samples
Save as User Program
To store the selected parameters as a User Program, refer to
Reading Mode
Single Reading: A single reading is displayed after Read is pressed.
Carousel 1 inch square: Optional measurement of carousel insert with up to 5 square cells.
Carousel 1 cm square: Optional measurement of carousel insert with up to 7 square cells.
Recall measurement
data
Recall saved measurement data, wavelength scans or time courses, see
.
Instrument Setup mode
Basic data of the instrument, refer to
.