Expanded programs

6.1.2.1

Measurement process

The measurement process defines the handling and the readings
of the test:

At which and how many wavelengths should readings be
completed?

How many absorbance readings must be completed?

When should the zero reading and the sample reading be
completed?

Are waiting times necessary between readings?

Should individual program sequences be repeated?

The elements of a measuring sequence, such as zero and
sample readings and the timers (reaction times, waiting times
etc.) are individually defined.

6.1.2.2

Enter a new measurement process

Note: Each component of the measurement process must be entered in
the order in which it will be completed.

[Z] key / Zeroing...

1. Press [Z] to program a zero reading.

2. Press Zeroing and enter the wavelength for which the zero

reading is to be performed.

3. If zero readings are to be carried out at a number of

wavelengths, repeat the steps starting with 1 for each
wavelength.

Note: The entered reading sequence is displayed.

Table 10 Definitions of the program points

Program point

Description

Name

Name of the analysis parameter

Version

An abbreviation or version number assigned by the user is entered here.

Measurement process

Exact definition of the test: the number of wavelengths at which readings are carried out, the
number of absorbance readings needed, the keys to be used, any waiting periods between
readings, etc.

Formula

Definition of the formulas with which the test result is calculated.

Variables

The number of variables shown in the display depends on the definition of the measurement
process and the formulas. Input of the numerical values of the wavelengths, factors, constants,
etc.

Timer 1, Timer 2,
Timer 3, Timer 4

This function can be used to enter abbreviations and defined times for up to four timers.
Highlight the corresponding line and press Edit. The timers are activated or deactivated with the
control boxes on the left of the display. In the next column, a selection can be made from a list of
names that describe the corresponding work step. In the third column, the times for each active
timer are entered.