Section 7
Maintenance
7.1
Cleaning requirements
7.1.1
Spectrophotometer
•
Only clean the housing, the cell compartments and all
accessories with a soft, damp cloth. A mild soap solution can
also be used.
•
Do not get excess water in the cell compartments.
•
Insert no brush and no sharp objects in the cell compartment
(1), to avoid damage to mechanical components.
•
Dry the cleaned parts carefully with a soft cotton cloth.
7.1.2
Display
•
Do not scratch the display. Never touch the display with ball
pens, pencils or similar pointed objects.
•
Clean the display with a soft, lint- and oil-free cotton cloth.
Diluted window cleaner liquid can also be used.
C A U T I O N
Potential Chemical, Biological Eye and Skin Hazards.
Only qualified personnel should conduct the tasks described in this
section of the manual.
N O T I C E
Remove all cells remaining in the instrument. Dispose of the cells or
their contents in an authorised manner.
D A N G E R
Chemical exposure hazard.
Gas from a chlorine compound and UV light reaction can cause death.
Do not use chlorine compounds for cleaning.
C A U T I O N
Potential Pinch, Eye, Burn and Chemical Hazards.
Before cleaning, always disconnect the instrument from the power
source.
N O T I C E
Never use solvents like turpentine, acetone or similar to clean the
instrument, including the display and accessories.