Troubleshooting

No valid data for these parameters!

Data analysis not possible, no
measurement data

Change the selection.

No valid data found!

View Data not possible in data log

Change the selection.

No help function present.

No measurement data present!

Data analysis settings cannot be
configured without measurement data.

Change the selection.

Control range not reached!

Data analysis limits not reached

This is a warning notice. The control
limit set was not reached.

Control range exceeded!

Data analysis limits exceeded.

This is a warning notice. The control
limit was exceeded.

Concentration too high!

Calculated concentration is higher than
999999

Dilute sample and measure again

Over measuring range

The measured absorbance is above the
calibration range of the test

Dilute sample and measure again

Under measuring range

The measured absorbance is below the
calibration range of the test

If possible, select a test with a lower
reading range or use a cuvette with a
longer path length

Possible interference by:

Interference Check

The analysis is possibly erroneous due
to interferences.

Possible interference from:

Interference Check

The analysis is possibly erroneous due
to interferences.

Next service is due!

Contact technical support for an
inspection of the instrument. For setting
the service reminder, refer to

section 6.7.7, page 130

Negative result!

The calculated result is negative

Check concentration of sample

Unstable lighting conditions!

Avoid direct sunlight at the measuring
location.

System check incorrect!

Measurement of air values failed

Switch the instrument off and then back
on again. If the system check is not
successful, contact technical support.

Temperature too high.
Measurement not possible!

Switch of the instrument and allow it to
cool for a few minutes. If necessary,
move it to a cooler place.

Update errors

An error occurred when uploading
the instrument data.

Start the procedure again or contact
technical support.

An error occurred when reading from
the USB memory stick.

Start the procedure again or contact
technical support.

An error occurred when writing to the
USB memory stick.

Start the procedure again or contact
technical support.

Please check on the current update
file.

Error during update.

Check the USB memory stick.

Please contact Customer Service.

Error during update.

Please insert the USB memory stick.

Insert a USB memory stick into a USB A
port on the instrument.

File for instrument update missing.

Error during update.

Check the USB memory stick.

Error displayed

Cause

Resolution